Abstract # 2216 Analysis of Interfractional Set-up Errors and Intrafractional Organ Motions during IMRT for Head and Neck Cancers toDefine an Appropriate PTV-margin

Presenter: Suzuki, Minoru

< Back to topics

> Cover Page, Authors, Institutions

> Purpose/Objective

> Materials/Methods

> Results

> Conclusions

> View tables/images/videos


Results:
The mean value for systematic interfractional setup errors of all the landmarks ranged within 1 mm along any directions. The standard deviation (SD) of the systematic set-up errors ( -INTER) distributed with a range of 0.7 to1.3 mm. The average of the SDs of the individual random interfractional set-up errors ( -INTER) ranged from 0.7 to 1.6 mm. The mean value for systematic interfractional organ motion errors of all the landmarks ranged within 0.5 mm along any directions. The SD of the systematic intrafractional organ motion errors ( -intra) distributed with a range of 0.2 to 0.8 mm. The average of the SD of the individual random intrafractional organ motion errors ( -intra) ranged from 0.3 to 0.6 mm. Using an equation of [PTV-margin = 2 + 0.7 ], the estimated appropriate PTV-margins for all the bony landmarks ranged from 2.0 to 3.6 mm.